Horizontal Rectangular Autoclave
Chandigarh, Punjab, India
Autoclave Steam Sterilizer - Medical and Laboratory Grade, Export from India
Shipping Information
Exporter details revealed after verified importer registration
The Horizontal Rectangular Autoclave (Model EI-R08) is a gravity displacement steam sterilizer built for bulk sterilization in hospitals, medical institutions, agricultural facilities, and pharmaceutical plants. It operates on the downward displacement of air principle - saturated steam at 121°C rapidly and thoroughly destroys microbiological agents across all types of bulk sterilization loads. Suitable for hospital dressings, linen, rubber and plastic goods, surgical instruments, glassware, and utensils. The chamber is constructed from Mild Steel or non-magnetic Stainless Steel 304/316 thick plate, all-weld construction that is leak-proof unlike riveted alternatives, and hydraulically tested to withstand 2.5 times the working pressure. India is among the top three global exporters of medical autoclaves alongside China and Italy, supplying verified buyers across 110+ countries including Uzbekistan, Tanzania, Sri Lanka, Nepal, and Mozambique.
| Attribute | Details / Value |
|---|---|
| Model | EI-R08 - Horizontal Rectangular Autoclave |
| Sterilization Principle | Downward (gravity) displacement of air using saturated steam |
| Sterilizing Temperature | 121°C |
| Sterilizing Pressure | 1.2 to 1.5 kg/cm² (15 to 22 PSI) |
| Chamber Material | Mild Steel / Stainless Steel 304/316 (non-magnetic thick plate) |
| Jacket Material | Mild Steel (outer side lined with asbestos sheet or glass wool for heat insulation) |
| Door | Single door, rolled steel with SS lining on both sides; Self Locking Safety Door |
| Boiler Material | Stainless Steel thick plate, located under the sterilizer |
| Heating Elements | High-grade flanged type, front-fitted for easy replacement and repair |
| Pressure Gauge | Individual gauges for chamber and boiler |
| Water Gauge Glass | Provided for water level monitoring |
| Low Water Level Cut Off | Provided - protects heater from burning out dry |
| Steam Trap | Provided |
| Ejector | Provided for drying sterilized linen |
| Switch Box | Pressure control switch, main switch, pilot lamps, relays |
| Construction | Robust all-weld, leak-proof (unlike riveted construction) |
| Testing | Hydraulically tested to withstand 2.5 times working pressure |
| HS Code | 8419.20 (medical, surgical or laboratory sterilizers) - confirm with CHA |
The EI-R08 runs a triple-layer safety system. First, a pressure control switch automatically limits chamber pressure to the required set value. Second, two spring-loaded safety valves - one each at chamber and boiler level - provide redundant pressure relief. Third, a dead weight safety valve operates at chamber level as a final failsafe. The door carries a self-locking mechanism that physically prevents opening while the chamber is under pressure - a hardware interlock, not a software prompt. For hospital procurement teams and biomedical engineers evaluating safety compliance, all three mechanisms are documented in the product test certificate supplied with each unit.
A plug screen prevents the discharge line from choking during operation. The plug is removable for routine cleaning without tools. A powerful ejector circulates air through the chamber for drying sterilized linen after the cycle. Circulating air passes through a corrosion-resistant metal wool filter. A dial thermometer shows live chamber temperature. An automatic vacuum breaker prevents vacuum formation caused by steam condensation at cycle end - a detail that matters for load integrity on wrapped and textile loads.
The following are available as optional upgrades at time of order: chamber and jacket upgrade to SS 316; door upgrade to SS 304/316; additional trays and shelves for increased load configuration; carriage and trolley for bulk load handling. Specify accessories at enquiry stage - retrofitting after delivery is not possible on chamber and jacket upgrades.
Hospital dressings, surgical linen, rubber goods, plastic items, surgical instruments, glassware, laboratory utensils, and general bulk loads in medical, agricultural, and pharmaceutical institutional settings. The downward displacement (gravity) cycle is the most economical method for bulk sterilization of these load types and is what the EI-R08 is designed and validated for. It is not a pre-vacuum B-class unit - buyers requiring hollow instrument or wrapped pack sterilization to EN 13060 B-class standards should specify that requirement separately.
Heavy-duty wooden crating with shock-absorbing mounts. Each unit ships with hydraulic test certificate, calibration documentation, user and service manual, spare parts kit (gaskets, heating elements, fuses, safety valve set), and installation guide. FCL container loading plans provided for multi-unit orders.
CE marking available for EU import, ISO 13485 (medical device quality management), hydraulic pressure test certificate (2.5x working pressure) issued per unit, BIS certification. Calibration and performance test documentation supplied with each unit. Additional compliance documentation for WHO-GMP, FDA registration, or country-specific medical device import requirements available on request.
The EI-R08 sterilizes at 121°C at a pressure of 1.2 to 1.5 kg/cm² (15 to 22 PSI). This is the standard gravity displacement cycle for bulk sterilization of hospital dressings, linen, rubber goods, surgical instruments, and glassware. The unit is hydraulically tested to withstand 2.5 times the working pressure before dispatch.
The EI-R08 is suitable for hospital dressings, surgical linen, rubber and plastic goods, surgical instruments, laboratory glassware, and general utensils. It operates on the downward displacement gravity principle, which is the most economical method for bulk sterilization loads in medical, agricultural, and pharmaceutical institutional settings. It is not designed for hollow instrument sterilization to B-class EN 13060 standards - if that is a requirement, specify it separately at enquiry stage.
The EI-R08 has a triple safety system: a pressure control switch that automatically limits pressure to the set value; two spring-loaded safety valves at chamber and boiler level; and a dead weight safety valve at chamber level. The door has a self-locking safety mechanism that cannot be opened while the chamber is under pressure. A low water level cut-off protects the heating elements from burning dry. An automatic vacuum breaker prevents vacuum formation from steam condensation at cycle end.
Optional upgrades include: SS 316 chamber and jacket (instead of standard MS/SS 304); SS 304/316 door; additional trays and shelves for increased load configuration; carriage and trolley for bulk load handling. All accessories and material upgrades must be specified at the time of order. Chamber and jacket material upgrades cannot be retrofitted after manufacture.
Each EI-R08 ships with a hydraulic test certificate (2.5x working pressure), calibration documentation, user and service manual, spare parts kit, and installation guide. CE marking and ISO 13485 documentation available for EU buyers. Standard lead time is 20-30 days. SS 316 chamber upgrades or large chamber configurations with full testing documentation run 40-55 days. Sea freight from Nhava Sheva to UAE/Gulf: 7-12 days; East Africa: 12-18 days; Europe: 20-25 days.
Connect with EXP01040 and thousands of verified Indian exporters on Navi Exports